Determination of SiO2、CaO、MgO、Al2O3、Fe2O3 in Diopside by Inductively Coupled Plasma Atomic Emission Spectrometry with Lithium Metaborate Fusion Sample Pretreatment
-
Graphical Abstract
-
Abstract
Lithium metaborate fusion pretreatment and inductively coupled plasma atomic emission spectrometry(ICP-AES) with internal standard method were used for the determination of SiO2、CaO、MgO、Al2O3、Fe2O3 in diopside. The best analytical lines of the elements were determined. The interferences were eliminated by using matched matrix and background correction. The experimental results showed a good linear relationship with the linear correlation coefficient(r ≥ 0.999 5), the detection limits of the elements were 0.061%(SiO2), 0.039%(CaO), 0.024%(MgO), 0.011%(Al2O3), 0.009%(Fe2O3), the precision of RSD were less than 5%, the recovery rates were in the range of 99.0%~101.8%. The proposed method was applied for the determination of real samples of diopside, and the results were basically consistent with those obtained by chemical analysis methods in other laboratories.
-
-