A New Chemiluminescence System for Determination of Copper
-
Graphical Abstract
-
Abstract
A new complex cheminescence, i.e. Luminol H2O2 catalyzed by CuR22- complex was reported. The linear range for the determination of Cu(Ⅱ) is 1×10-10~1×10-7g/mL. The detection limit is 6.6×10-11 g/mL. The relative standard deviation (n=5) for an analytical concentration of 4×10-8 g/mL is 2.7%. The tolerable capacity of interserence ion is relatively large.The precision, accuracy and selectivity of the method are very good.The mothod has been used to determine trace copper with satisactory results.
-
-