FENG Shi-lan, MENG Yan, HU Fang-di, CHEN Li-ren. Determination of Phenolic Glycosides from Salix Oritrepha by High-performance Liquid Chromatography[J]. Analysis and Testing Technology and Instruments, 2002, (4): 236-239.
Citation: FENG Shi-lan, MENG Yan, HU Fang-di, CHEN Li-ren. Determination of Phenolic Glycosides from Salix Oritrepha by High-performance Liquid Chromatography[J]. Analysis and Testing Technology and Instruments, 2002, (4): 236-239.

Determination of Phenolic Glycosides from Salix Oritrepha by High-performance Liquid Chromatography

  • Salicin, Salicyloylsalicin, and Leteolin 7-O-glucoside were determined from S. Oritrepha. Methods: High-performance liquid chromatography(HPLC) on a Symmetry C18 column (4.6 mm×250 mm, 5 μm) with acetic acid-methanol gradient elution, at column temperature of 30℃, flow-rate of 1.1 mL/min and wavelength at 270 nm. Results: The contents of three compounds can be determined simultaneously from S. Oritrepha.The linear concentration ranges of these compounds were 527.0~105.4 μg/mL, 140.0~28.0 μg/mL, 81.0~16.2 μg/mL respectively. The average recoveries were 104.9%, 104.6% and 103.6% respectively, and relative standard deviations were 2.0%, 0.3% and 0.3% respectively. Conclusion: The proposed method was a simple and rapid method for simultaneous determination of the three mentioned compounds in various tissues from S. Oriterpha.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return